Mass fractionation of Rb and Sr isotopes during laser ablation-multicollector-ICPMS: in situ observation and correction
Background One of the most critical issues concerning in situ mass spectrometry lies in accounting for elements and molecules that overlap target isotopes of analytical interest in a sample. This study traced the instrumental mass fractionation of Rb and Sr isotopes during laser ablation-multicollec...
Gespeichert in:
Veröffentlicht in: | Journal of Analytical Science and Technology 2021-02, Vol.12 (1), p.1-10, Article 10 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background
One of the most critical issues concerning in situ mass spectrometry lies in accounting for elements and molecules that overlap target isotopes of analytical interest in a sample. This study traced the instrumental mass fractionation of Rb and Sr isotopes during laser ablation-multicollector-inductively coupled plasma mass spectrometry (LA-MC-ICPMS) to obtain reliable
87
Sr/
86
Sr ratios for high-Rb/Sr samples.
Findings
In the LA-MC-ICPMS analysis, Kr interferences were corrected using Ar and He gas blanks measured without ablating material. Contributions from doubly charged Er and Yb ions were corrected using the intensities of half masses and isotopic compositions reported in the literature. After Kr correction, the calculated
166
Er
2+
intensity of NIST SRM 610 approached the measured intensity at mass 83, and the
173
Yb
2+
/
171
Yb
2+
ratio agreed with the recommended value within error ranges. Kr- and REE
2+
-stripped peak intensities were further corrected for Rb interference. Use of the Sr mass bias factor for the calculation of measured
87
Rb/
85
Rb yielded
87
Sr/
86
Sr ratios consistent with the recommended and expected values for low-Rb/Sr materials, such as NIST SRM 616, modern shark teeth, and plagioclase collected from Jeju Island, but failed to account for the
87
Rb interference from high-Rb/Sr materials including NIST SRM 610 and SRM 612. We calculated in situ mass bias factor of Rb from the known
87
Sr/
86
Sr ratios of the standards and observed a correlation between Rb and Sr mass fractionation, which allowed inference of the Rb bias from the standard run. Reliable
87
Sr/
86
Sr and
85
Rb/
86
Sr ratios were obtained for SRM 610 and SRM 612 using the inferred mass bias factor of Rb calculated by the standard bracketing method.
Conclusions
This study revealed that Rb and Sr isotopes behave differently during LA-MC-ICPMS and suggests the potential usefulness of the standard bracketing method for measuring the Rb–Sr isotopic compositions of high-Rb/Sr materials. |
---|---|
ISSN: | 2093-3371 2093-3134 2093-3371 |
DOI: | 10.1186/s40543-021-00263-9 |