Effect of initial switch-on within 24 hours of cochlear implantation using slim modiolar electrodes
Reducing electrode impedance is an important factor in improving the functional benefits of cochlear implants (CIs). The immediate effect of early switch-on within 24 h of surgery on impedance among CI recipients with various types of electrodes has been reported previously; however, the immediate c...
Gespeichert in:
Veröffentlicht in: | Scientific reports 2021-11, Vol.11 (1), p.22809-22809, Article 22809 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Reducing electrode impedance is an important factor in improving the functional benefits of cochlear implants (CIs). The immediate effect of early switch-on within 24 h of surgery on impedance among CI recipients with various types of electrodes has been reported previously; however, the immediate change and the evolution of electrode impedances of slim modiolar electrodes after early switch-on within 24 h of implantation has not. Therefore, the focus of this retrospective cohort study of CI patients was to compare the effect of early switch-on (n = 36) and conventional switch-on (n = 72) 2–4 weeks post-operation on impedance. Compared with impedance measured intraoperatively, our results demonstrate a significant decrease in impedance from 11.5 to 8.9 kΩ (
p
|
---|---|
ISSN: | 2045-2322 2045-2322 |
DOI: | 10.1038/s41598-021-01862-7 |