Effectiveness of the C-MAC video laryngoscope in the management of unexpected failed intubations
The purpose of this study was to review the experiences of an anesthesiology department regarding the use of a C-MAC videolaryngoscope in unexpected failed intubation attempts. Data were analyzed from 42 patients whose intubation attempts using Macintosh direct laryngoscopes had failed, and on whom...
Gespeichert in:
Veröffentlicht in: | Brazilian journal of anesthesiology (Elsevier) 2014-01, Vol.64 (1), p.62-65 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The purpose of this study was to review the experiences of an anesthesiology department regarding the use of a C-MAC videolaryngoscope in unexpected failed intubation attempts.
Data were analyzed from 42 patients whose intubation attempts using Macintosh direct laryngoscopes had failed, and on whom a C-MAC videolaryngoscope was utilized as the primary rescue device. The success rate of C-MAC in intubation was assessed, and laryngeal views from both devices were compared.
The Cormack and Lehane score was III in 41 patients, and IV in one patient, with the Macintosh laryngoscope, while Cormack and Lehane score was I in 27 patients, II in 14 and III in one with CMAC. Tracheal intubation with CMAC was successful on the first attempt in 36 patients (86%), and on the second attempt in 6 patients (14%). No complications were observed other than minor damage (blood on blade) in 8 patients (19%).
These data provide evidence for the clinical effectiveness of C-MAC videolaryngoscope in managing the unexpected failed intubations in routine anesthesia care. The C-MAC videolaryngoscope is efficient and safe as a primary rescue device in unexpected failed intubations. |
---|---|