Radiotherapy planning of spine and pelvis using single-energy metal artifact reduction corrected computed tomography sets

Metal artifacts produce incorrect Hounsfield units and impact treatment planning accuracy. This work evaluates the use of single-energy metal artifact reduction (SEMAR) algorithm for treatment planning by comparison to manual artifact overriding. CT datasets of in-house 3D-printed spine and pelvic p...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Physics and imaging in radiation oncology 2023-04, Vol.26, p.100449-100449, Article 100449
Hauptverfasser: Ignatius, Daliya, Alkhatib, Zaid, Rowshanfarzad, Pejman, Goodall, Simon, Ibrahim, Mounir, Hirst, Andrew, Croxford, Riley, Dass, Joshua, Sabet, Mahsheed
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Metal artifacts produce incorrect Hounsfield units and impact treatment planning accuracy. This work evaluates the use of single-energy metal artifact reduction (SEMAR) algorithm for treatment planning by comparison to manual artifact overriding. CT datasets of in-house 3D-printed spine and pelvic phantoms with and without metal insert(s) and two treated patients with metal implants were analysed. CT number accuracy improved with the use of SEMAR filter: root mean square deviation (RMSD) from reference (without metal) reduced by 35.4 in spine and 98.8 in hip. The plan dose volume histograms (DVHs) and dosimetric measurements showed comparable results. SEMAR reconstruction improved planning efficiency.