An automated dose verification software for brachytherapy
To report an implementation method and the results of independent brachytherapy dose verification software (DVS). The DVS was developed based on Visual C++ and adopted a modular structure design. The DICOM RT files exported from a treatment planning system (TPS) were automatically loaded into the DV...
Gespeichert in:
Veröffentlicht in: | Journal of contemporary brachytherapy 2018-10, Vol.10 (5), p.478-482 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To report an implementation method and the results of independent brachytherapy dose verification software (DVS).
The DVS was developed based on Visual C++ and adopted a modular structure design. The DICOM RT files exported from a treatment planning system (TPS) were automatically loaded into the DVS. The DVS used the TG-43 formalism for dose calculation. A total of 15 cervical cancer patients who underwent brachytherapy were retrospectively selected to test the DVS. Dosimetric parameters and γ analysis (0.1 cm, 5%) were used to evaluate the dose differences between the DVS and the TPS.
Compared with the TPS dose, the γ pass rates of the dose calculated by the DVS were higher than 98%. For the clinical target volume (CTV), the dosimetric differences were less than 0.63% for D
and D
. For the bladder, rectum, and sigmoid, the agreement of D
, D
, and D
were within a 0.78% level.
With minimal human-computer interactions, the DVS can verify the accuracy of doses calculated by the TPS. |
---|---|