Performance of direct immunofluorescence assay for the detection of human metapneumovirus under clinical laboratory settings

Human metapneumovirus (hMPV) is an emergent human respiratory pathogen. This study aimed to evaluate the performance of direct immunofluorescence (DIF) to detect hMPV in a clinical laboratory setting. Nasopharyngeal aspirate samples (448) of children and adults with respiratory illness were used to...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Revista da Sociedade Brasileira de Medicina Tropical 2015-12, Vol.48 (6), p.762-764
Hauptverfasser: Wolf, Jonas Michel, Gregianini, Tatiana Schäffer, Seadi, Claudete Maria Farina, Tumioto, Gabriela Luchiari, Dambrós, Bibiana Paula, Lehmann, Fernanda Kieling Moreira, Carli, Silvia De, Ikuta, Nilo, Lunge, Vagner Ricardo
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Human metapneumovirus (hMPV) is an emergent human respiratory pathogen. This study aimed to evaluate the performance of direct immunofluorescence (DIF) to detect hMPV in a clinical laboratory setting. Nasopharyngeal aspirate samples (448) of children and adults with respiratory illness were used to detect hMPV by using DIF and real time quantitative reverse transcription-polymerase chain reaction (qRT-PCR) assays. In all, 36 (8%) samples were positive by DIF and 94 (21%) were positive by qRT-PCR. Direct immunofluorescence specificity was 99% and sensitivity was 38%. DIF is not very sensitive under clinical laboratory settings.
ISSN:0037-8682
1678-9849
1678-9849
0037-8682
DOI:10.1590/0037-8682-0107-2015