Electrostatic Discharge Causing Pump Shutdown in HeartMate 3

Left ventricular assist devices (LVADs) improve symptoms and outcomes in advanced heart failure. Although device malfunction has decreased significantly with later generation LVADs, it has not been eliminated. We describe the clinical course of a patient with HeartMate 3 LVAD who experienced device...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:JACC. Case reports 2021-03, Vol.3 (3), p.459-463
Hauptverfasser: Najjar, Emil, Hallberg Kristensen, Ann, Thorvaldsen, Tonje, Dalén, Magnus, Jorde, Ulrich P., Lund, Lars H.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Left ventricular assist devices (LVADs) improve symptoms and outcomes in advanced heart failure. Although device malfunction has decreased significantly with later generation LVADs, it has not been eliminated. We describe the clinical course of a patient with HeartMate 3 LVAD who experienced device malfunction, involving temporary pump shutdown suspected to be caused by electrostatic discharge. ( Level of Difficulty: Advanced. )
ISSN:2666-0849
2666-0849
DOI:10.1016/j.jaccas.2020.12.036