Assessing the accuracy of photoplethysmography for wearable heart rate monitoring based on body location and body motion in uncontrolled outdoor environments

Introduction Reflective photoplethysmography (PPG) is the dominant method for heart rate (HR) monitoring consumer wearables. However, motion artifacts and sensor placement impact the accuracy of these HR measurements. Here, we present a study on how these two factors affect the accuracy of PPG-based...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Current Issues in Sport Science 2024-02, Vol.9 (2), p.59
Hauptverfasser: Meier, Manuel, Holz, Christian
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Introduction Reflective photoplethysmography (PPG) is the dominant method for heart rate (HR) monitoring consumer wearables. However, motion artifacts and sensor placement impact the accuracy of these HR measurements. Here, we present a study on how these two factors affect the accuracy of PPG-based HR measurements and compare them to ground-truth measurements from electrocardiography (ECG). Our study investigated these measurements in outdoor environments outside controlled laboratory settings. Methods Our study collected a dataset of 16 participants, each for 13 hours wearing four reflective PPG sensing devices placed at the forehead, sternum, ankle (supramalleolar), and wrist. Participants traveled by train from downtown Zurich to the Jungfraujoch railway station at 3,460 m above sea level in the mountains. PPG measurements were obtained using a MAX86141 optical analog front-end coupled with a green LED and photodiode from an SFH7072 module. Motion was quantified using two accelerometers (LIS2DH, ADXL355). A Lead I ECG was obtained by the device at the sternum using a MAX30003 biopotential sensor connected to gel electrodes on the chest. All devices were synchronized by aligning recorded signals post-hoc (33ms accuracy, Meier & Holz, 2023). The HR was extracted from the ECG by time-domain peak detection. The HR extraction from PPG was both performed by time-domain peak detection and frequency-domain analysis. The HR was computed every 5 seconds (30 seconds window size) resulting in 152,000 HR measurements across the whole dataset. Results The forehead and chest locations exhibited the highest HR measurement accuracy (median error 7.1% and 7.7%, respectively), while lower accuracies were observed for ankle and wrist placements (9.9% and 18.4% error). At rest, all median errors were below 5% while movements influenced readings at all locations negatively. Adjusted for motion, the HR obtained from the forehead sensor was most accurate. In terms of processing method, time-domain analysis produced better accuracy during periods of low motion while frequency-domain analysis was more reliable during movements. Discussion/Conclusion The accuracy of PPG-based HR measurements in uncontrolled outdoor settings is affected both by body location and motion artifacts with a clear ranking of site suitability: forehead >> chest >> ankle >> wrist. This is consistent with prior studies in controlled environments, though our study found a higher impact of motion than body
ISSN:2414-6641
2414-6641
DOI:10.36950/2024.2ciss059