COVID severity test ( CoST sensor)—An electrochemical immunosensing approach to stratify disease severity
With the evolution of the COVID-19 pandemic, there is now a need for point-of-care devices for the quantification of disease biomarkers toward disease severity assessment. Disease progression has been determined as a multifactor phenomenon and can be treated based on the host immune response within...
Gespeichert in:
Veröffentlicht in: | Bioengineering & Translational Medicine 2023-09, Vol.8 (5), p.e10566-e10566 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | With the evolution of the COVID-19 pandemic, there is now a need for point-of-care devices for the quantification of disease biomarkers toward disease severity assessment. Disease progression has been determined as a multifactor phenomenon and can be treated based on the host immune response within each individual. CoST is an electrochemical immunosensor point-of-care device that can determine disease severity through multiplex measurement and quantification of spike protein, nucleocapsid protein, D-dimer, and IL-2R from 100 μL of plasma samples within a few minutes. The limit of detection was found to be 3 ng/mL and 21 ng/mL for S and N proteins whereas for D-dimer and IL-2R it was 0.0006 ng/mL and 0.242 ng/mL, respectively. Cross-reactivity of all the biomarkers was studied and it was found to be |
---|---|
ISSN: | 2380-6761 2380-6761 |
DOI: | 10.1002/btm2.10566 |