Simulation of the 6 MV Elekta Synergy Platform linac photon beam using Geant4 Application for Tomographic Emission
The present work validates the Geant4 Application for Tomographic Emission Monte Carlo software for the simulation of a 6 MV photon beam given by Elekta Synergy Platform medical linear accelerator treatment head. The simulation includes the major components of the linear accelerator (LINAC) with mul...
Gespeichert in:
Veröffentlicht in: | Journal of medical physics 2015-07, Vol.40 (3), p.136-143 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The present work validates the Geant4 Application for Tomographic Emission Monte Carlo software for the simulation of a 6 MV photon beam given by Elekta Synergy Platform medical linear accelerator treatment head. The simulation includes the major components of the linear accelerator (LINAC) with multi-leaf collimator and a homogeneous water phantom. Calculations were performed for the photon beam with several treatment field sizes ranging from 5 cm × 5 cm to 30 cm × 30 cm at 100 cm distance from the source. The simulation was successfully validated by comparison with experimental distributions. Good agreement between simulations and measurements was observed, with dose differences of about 0.02% and 2.5% for depth doses and lateral dose profiles, respectively. This agreement was also emphasized by the Kolmogorov-Smirnov goodness-of-fit test and by the gamma-index comparisons where more than 99% of the points for all simulations fulfill the quality assurance criteria of 2 mm/2%. |
---|---|
ISSN: | 0971-6203 1998-3913 |
DOI: | 10.4103/0971-6203.165077 |