Left ventricular assist device implantation outcomes in patients with subcutaneous implantable cardioverter-defibrillators: A case series
Left ventricular assist devices (LVADs) may induce electromagnetic interference (EMI) affecting implanted cardiac devices, including more novel subcutaneous implantable cardiac defibrillators (S-ICDs). In this case series, the authors retrospectively reviewed courses of 6 patients with S-ICDs who un...
Gespeichert in:
Veröffentlicht in: | American heart journal plus 2024-09, Vol.45, p.100426, Article 100426 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Left ventricular assist devices (LVADs) may induce electromagnetic interference (EMI) affecting implanted cardiac devices, including more novel subcutaneous implantable cardiac defibrillators (S-ICDs).
In this case series, the authors retrospectively reviewed courses of 6 patients with S-ICDs who underwent LVAD implantation at a single center.
Of the 6 patients reviewed, 4 experienced inappropriate ICD shocks, of which 3 resulted from EMI. Five of the 6 patients ultimately had S-ICD therapies disabled.
Due to EMI resulting in inappropriate shocks and improved tolerability of malignant arrhythmias, deactivation or removal of S-ICDs should be considered in patients undergoing LVAD implantation. |
---|---|
ISSN: | 2666-6022 2666-6022 |