PRECICE ® magnetically-driven, telescopic, intramedullary lengthening nail: pre-clinical testing and first 30 patients

Femoral/tibial lengthening with a telescopic, magnetically-powered, intramedullary nail is an alternative to lengthening with external fixation. Pre-clinical testing was conducted of the PRECICE in a human cadaver. A retrospective review of the first 30 consecutive patients who underwent unilateral...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:SICOT-J 2017, Vol.3, p.19-19
Hauptverfasser: Wagner, Pablo, Burghardt, Rolf D, Green, Stuart A, Specht, Stacy C, Standard, Shawn C, Herzenberg, John E
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Femoral/tibial lengthening with a telescopic, magnetically-powered, intramedullary nail is an alternative to lengthening with external fixation. Pre-clinical testing was conducted of the PRECICE in a human cadaver. A retrospective review of the first 30 consecutive patients who underwent unilateral lengthening was also conducted. Nail accuracy was obtained by comparing the amount of nail distraction to the final bone length achieved at the end of the distraction process. Relative standard deviation of accuracy was used to calculate nail precision. Devices performed successfully in a human cadaver. Thirty consecutive patients (10 females, 20 males; mean age, 23 years) with limb length discrepancy (LLD) were followed an average of 19 months (range, 12-24 months). Etiology included congenital shortening (14), posttraumatic deformities (7), Ollier disease (3), osteosarcoma resection (1), prior clubfoot (2), hip dysplasia (1), post-septic growth arrest of knee (1), and LLD after hip arthroplasty (1). Twenty-four femoral and eight tibial nails were implanted. Mean preoperative lengthening goal was 4.4 cm (range, 2-6.5 cm); mean postoperative length achieved was 4.3 cm (range, 1.5-6.5 cm). Average consolidation index was 36.4 days/cm (range, 12.8-113 days/cm). Mean nail accuracy was 97.3% with a precision of 92.4%. Average preoperative and 12-month postoperative Enneking scores were 21.5 and 25.3 (p 
ISSN:2426-8887
2426-8887
DOI:10.1051/sicotj/2016048