Trends in maintenance status and usability of public automated external defibrillators during a 5-year on-site inspection
This study aimed to assess the trend of the maintenance status and usability of public automated external defibrillators (AEDs). Public AEDs installed in Seoul from 2013 to 2017 were included. An inspector checked the maintenance status and usability of the AEDs annually using a checklist. During th...
Gespeichert in:
Veröffentlicht in: | Scientific reports 2022-06, Vol.12 (1), p.10738-10738, Article 10738 |
---|---|
Hauptverfasser: | , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This study aimed to assess the trend of the maintenance status and usability of public automated external defibrillators (AEDs). Public AEDs installed in Seoul from 2013 to 2017 were included. An inspector checked the maintenance status and usability of the AEDs annually using a checklist. During the study period, 23,619 AEDs were inspected. Access to the AEDs was improved, including the absence of obstacles near the AEDs (from 90.2% in 2013 to 99.1% in 2017, p |
---|---|
ISSN: | 2045-2322 2045-2322 |
DOI: | 10.1038/s41598-022-14611-1 |