A Novel Technique for Identification of Wear Values at Different Lengths after Multiple Clinical Use of Different File Systems

Introduction The purpose of this research is to assess the wearing of the rotary file system (Protaper Next) and reciprocating file systems (Reciproc Blue and WaveOne Gold) at different lengths using a novel technique after in vivo clinical use. Materials and Methods Twelve different unused samples...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medicina (Kaunas, Lithuania) Lithuania), 2022-08, Vol.58 (8), p.1117
Hauptverfasser: Çırakoğlu, Neslihan Yılmaz, Çiçek, Ersan, Özarpa, Cevat, Özbay, Yağız, Özdemir, Olcay
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Introduction The purpose of this research is to assess the wearing of the rotary file system (Protaper Next) and reciprocating file systems (Reciproc Blue and WaveOne Gold) at different lengths using a novel technique after in vivo clinical use. Materials and Methods Twelve different unused samples from each brand were accepted as reference values. For three different brands, the diameters of the files were measured by taking 12 samples used once, 12 samples used twice, and 12 samples used three times. Images were taken with a USB Micron Microscope, and file diameters were measured by determining limit values with Autocad. Result Reciproc Blue system was the most worn at apical 1 mm, and WaveOne Gold system was the most worn at apical 3 mm. PTN system exhibited the least wearing at any length. Moreover, less wearing was observed in the rotation motion than in the reciprocating motion. Conclusion In clinical practice, for the guttapercha to be fully adapted to the apical construction prepared according to the determined WL, the file should not undergo any wearing and volume reduction. Wearing—especially in the apical parts of the file—causes less preparation, and this situation could lead to apically obturation failure.
ISSN:1648-9144
1010-660X
1648-9144
DOI:10.3390/medicina58081117