Transient Asystole From Pacing Inhibition During Percutaneous Coronary Intervention

A 68-year-old man with a biventricular implantable cardioverter-defibrillator who was pacemaker dependent was admitted for percutaneous coronary intervention. The patient had an asystolic cardiac arrest as a result of electromagnetic interference during an orbital atherectomy. We should keep electro...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:JACC. Case reports 2021-01, Vol.3 (1), p.144-145
Hauptverfasser: Ishizu, Kenichi, Isotani, Akihiro, Hyodo, Makoto, Shirai, Shinichi, Ando, Kenji
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A 68-year-old man with a biventricular implantable cardioverter-defibrillator who was pacemaker dependent was admitted for percutaneous coronary intervention. The patient had an asystolic cardiac arrest as a result of electromagnetic interference during an orbital atherectomy. We should keep electromagnetic interference from the machine in mind when performing orbital atherectomy. (Level of Difficulty: Beginner.) [Display omitted]
ISSN:2666-0849
2666-0849
DOI:10.1016/j.jaccas.2020.10.029