Influence of Different Patellofemoral Design Variations Based on Genesis II Total Knee Endoprosthesis on Patellofemoral Pressure and Kinematics
In total knee arthroplasty (TKA), patellofemoral groove design varies greatly and likely has a distinct influence on patellofemoral biomechanics. To analyse the selective influence, five patellofemoral design variations were developed based on Genesis II total knee endoprosthesis (original design, b...
Gespeichert in:
Veröffentlicht in: | Applied bionics and biomechanics 2017-01, Vol.2017 (2017), p.1-10 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In total knee arthroplasty (TKA), patellofemoral groove design varies greatly and likely has a distinct influence on patellofemoral biomechanics. To analyse the selective influence, five patellofemoral design variations were developed based on Genesis II total knee endoprosthesis (original design, being completely flat, being laterally elevated, being medially elevated, and both sides elevated) and made from polyamide using rapid prototyping. Muscle-loaded knee flexion was simulated on 10 human knee specimens using a custom-made knee simulator, measuring the patellofemoral pressure distribution and tibiofemoral and patellofemoral kinematics. The measurements were carried out in the native knee as well as after TKA with the 5 design prototypes. The overall influence of the different designs on the patellofemoral kinematics was small, but we found detectable effects for mediolateral tilt (p |
---|---|
ISSN: | 1176-2322 1754-2103 |