Non-invasive estimation of hemoglobin, bilirubin and oxygen saturation of neonates simultaneously using whole optical spectrum analysis at point of care

The study was aimed to evaluate the performance of a newly developed spectroscopy-based non-invasive and noncontact device (SAMIRA) for the simultaneous measurement of hemoglobin, bilirubin and oxygen saturation as an alternative to the invasive biochemical method of blood sampling. The accuracy of...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Scientific reports 2023-02, Vol.13 (1), p.2370-2370, Article 2370
Hauptverfasser: Banerjee, Amrita, Bhattacharyya, Neha, Ghosh, Ria, Singh, Soumendra, Adhikari, Aniruddha, Mondal, Susmita, Roy, Lopamudra, Bajaj, Annie, Ghosh, Nilanjana, Bhushan, Aman, Goswami, Mahasweta, Ahmed, Ahmed S. A., Moussa, Ziad, Mondal, Pulak, Mukhopadhyay, Subhadipta, Bhattacharyya, Debasis, Chattopadhyay, Arpita, Ahmed, Saleh A., Mallick, Asim Kumar, Pal, Samir Kumar
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The study was aimed to evaluate the performance of a newly developed spectroscopy-based non-invasive and noncontact device (SAMIRA) for the simultaneous measurement of hemoglobin, bilirubin and oxygen saturation as an alternative to the invasive biochemical method of blood sampling. The accuracy of the device was assessed in 4318 neonates having incidences of either anemia, jaundice, or hypoxia. Transcutaneous bilirubin, hemoglobin and blood saturation values were obtained by the newly developed instrument which was corroborated with the biochemical blood tests by expert clinicians. The instrument is trained using Artificial Neural Network Analysis to increase the acceptability of the data. The artificial intelligence incorporated within the instrument determines the disease condition of the neonate. The Pearson’s correlation coefficient, r was found to be 0.987 for hemoglobin estimation and 0.988 for bilirubin and blood gas saturation respectively. The bias and the limits of agreement for the measurement of all the three parameters were within the clinically acceptance limit.
ISSN:2045-2322
2045-2322
DOI:10.1038/s41598-023-29041-w