Considerations in the use of different spirometers in epidemiological studies

Spirometric lung function measurements have been proven to be excellent objective markers of respiratory morbidity. The use of different types of spirometers in epidemiological and clinical studies may present systematically different results affecting interpretation and implication of results. We a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Environmental health 2019-04, Vol.18 (1), p.39-39, Article 39
Hauptverfasser: Milanzi, Edith B, Koppelman, Gerard H, Oldenwening, Marieke, Augustijn, Sonja, Aalders-de Ruijter, Bernadette, Farenhorst, Martijn, Vonk, Judith M, Tewis, Marjan, Brunekreef, Bert, Gehring, Ulrike
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Spirometric lung function measurements have been proven to be excellent objective markers of respiratory morbidity. The use of different types of spirometers in epidemiological and clinical studies may present systematically different results affecting interpretation and implication of results. We aimed to explore considerations in the use of different spirometers in epidemiological studies by comparing forced expiratory volume in 1 s (FEV ) and forced vital capacity (FVC) measurements between the Masterscreen pneumotachograph and EasyOne spirometers. We also provide a correction equation for correcting systematic differences using regression calibration. Forty-nine volunteers had lung function measured on two different spirometers in random order with at least three attempts on each spirometer. Data were analysed using correlation plots, Bland and Altman plots and formal paired t-tests. We used regression calibration to provide a correction equation. The mean (SD) FEV and FVC was 3.78 (0.63) L and 4.78 (0.63) L for the Masterscreen pneumotachograph and 3.54 (0.60) L and 4.41 (0.83) L for the EasyOne spirometer. The mean FEV difference of 0.24 L and mean FVC difference of 0.37 L between the spirometers (corresponding to 6.3 and 8.4% difference, respectively) were statistically significant and consistent between younger ( 30 years) and between males and females. Regression calibration indicated that an increase of 1 L in the EasyOne measurements corresponded to an average increase of 1.032 L in FEV and 1.005 L in FVC in the Masterscreen measurements. Use of different types of spirometers may result in significant systematic differences in lung function values. Epidemiological researchers need to be aware of these potential systematic differences and correct for them in analyses using methods such as regression calibration.
ISSN:1476-069X
1476-069X
DOI:10.1186/s12940-019-0478-2