Electrochemical Detection of Pb(II) Using A Pencil Electrode with Square Wave Anodic Stripping Voltammetry Method
This work describes a simple and inexpensive Pb(II) determination by a pencil electrode using the square wave anodic stripping voltammetry (SWASV) method. This research was applied using a batch cell system on 10 mL of 0.1 M acetate buffer pH 4.5. The precision was performed with repeatability and r...
Gespeichert in:
Veröffentlicht in: | EduChemia 2022-07, Vol.7 (2), p.150-162 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng ; ind |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This work describes a simple and inexpensive Pb(II) determination by a pencil electrode using the square wave anodic stripping voltammetry (SWASV) method. This research was applied using a batch cell system on 10 mL of 0.1 M acetate buffer pH 4.5. The precision was performed with repeatability and reproducibility in Pb(II) concentrations of 0. 2, 0. 4, 0. 6, and 0.8 µg mL-1 . Two linear ranges were obtained in the concentration of 0.1–1.0 and 2.0–8.0 µg mL-1. This method offers a detection limit of 0.07 µg mL-1 and a quantification limit of 0.24 µg mL-1. The technique performed good repeatability (RSD= 7.28–9.53%) and good reproducibility (RSD=7.07–15.10%). The precisions are accepted according to the Association of Official Analytical Chemists (AOAC) standard (RSD repeatability |
---|---|
ISSN: | 2502-4779 2502-4787 |
DOI: | 10.30870/educhemia.v7i2.14756 |