Unexplained rupture after endovascular aneurysm repair

Abstract We present a case of a 70-year-old man who was admitted with rupture of an abdominal aneurysm 4 years after endovascular aneurysm repair. He was compliant with yearly follow-up computed tomography angiography. One month earlier, his computed tomography angiogram showed perfect exclusion of...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of vascular surgery cases and innovative techniques 2017-09, Vol.3 (3), p.126-128
Hauptverfasser: van Rijn, Marie Josee Elizabeth, MD, PhD, ten Raa, Sander, MD, PhD, Hendriks, Joke M., MD, PhD, Bastos Goncalves, Frederico, MD, PhD, Verhagen, Hence J.M., MD, PhD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract We present a case of a 70-year-old man who was admitted with rupture of an abdominal aneurysm 4 years after endovascular aneurysm repair. He was compliant with yearly follow-up computed tomography angiography. One month earlier, his computed tomography angiogram showed perfect exclusion of the aneurysm and no endoleak. We explanted the stent graft and confirmed effective sealing, and the graft was intact. We found no signs of infection during 2 years of follow-up. This rupture is nonpredictable and unexplained and illustrates that unremarkable imaging does not guarantee prevention of rupture. This case shows that the ultimate failure of endovascular aneurysm repair cannot be prevented despite surveillance protocols.