Comparison between Roche Integra 400 plus and Abbott Architect ci8200 in Alanine aminotransferase assay
The aim of this work is to present the results of a comparative study between the ALT assay on Integra 400 plus Roche Diagnostic versus Architect ci8200 of Abbott Diagnostic. A total of 200 patients hospitalized in the various departments of the university hospital Mohammed VI of Oujda were prospect...
Gespeichert in:
Veröffentlicht in: | Practical laboratory medicine 2019-05, Vol.15, p.e00121-e00121, Article e00121 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The aim of this work is to present the results of a comparative study between the ALT assay on Integra 400 plus Roche Diagnostic versus Architect ci8200 of Abbott Diagnostic.
A total of 200 patients hospitalized in the various departments of the university hospital Mohammed VI of Oujda were prospectively tested on two systems: Abbott Architect ci8200 and Roche Integra 400 plus. Both analyzers use the spectrophotometric technique by coupling the transamination reaction to the oxidation-reduction reaction at NAD. The agreement of the results between the different techniques was evaluated using the Bland-Altman difference diagram and the Passing-Bablok and Deming regression line.
There was a high concordance between the two assays: the equation of the Passing-Bablok line is YArchitect = −0,5625 + 0,9917 XIntegra with a correlation coefficient r2 = 0.999. The Bland-Altman diagram shows that the mean bias between the two methods is of the order of 1.1 IU/L and the difference between the ALT measurements by Architect and Integra is in the range of −1.4 to 3.6.
Our study shows a high correlation of the ALT assay results between the architect ci8200 and Integra 400 plus. |
---|---|