Stability Indicating RP‐HPLC Method for Determination of Valsartan in Pure and Pharmaceutical Formulation
A simple, rapid and accurate and stability indicating RP‐HPLC method was developed for the determination of valsartan in pure and tablet forms. The method showed a linear response for concentrations in the range of 50‐175 µg/mL using 0.01 M NH 4 H 2 PO 4 (pH 3.5) buffer: methanol [50:50] as the mobi...
Gespeichert in:
Veröffentlicht in: | E-journal of chemistry 2010-01, Vol.7 (1), p.246-252 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A simple, rapid and accurate and stability indicating RP‐HPLC method was developed for the determination of valsartan in pure and tablet forms. The method showed a linear response for concentrations in the range of 50‐175 µg/mL using 0.01 M NH 4 H 2 PO 4 (pH 3.5) buffer: methanol [50:50] as the mobile phase with detection at 210 nm and a flow rate of 1 mL/min and retention time 11.041 min. The method was statistically validated for accuracy, precision, linearity, ruggedness, robustness, forced degradation, solution stability and selectivity. Quantitative and recovery studies of the dosage form were also carried out and analyzed; the % RSD from recovery studies was found to be less than 1. Due to simplicity, rapidity and accuracy of the method, we believe that the method will be useful for routine quality control analysis. |
---|---|
ISSN: | 2090-9063 0973-4945 2090-9071 2090-9810 |
DOI: | 10.1155/2010/487197 |