Accuracy of pilot balloon palpation for cuff pressure assessment in small versus large sized tubes: a prospective non-randomized observational study
Pilot balloon palpation is still a commonly used method to evaluate cuff pressure of the endotracheal tube after intubation. This study determined whether the size of the tracheal tube influenced the accuracy of pilot balloon palpation. A prospective observational analysis of 208 patients intubated...
Gespeichert in:
Veröffentlicht in: | Scientific reports 2023-04, Vol.13 (1), p.5580-5580, Article 5580 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Pilot balloon palpation is still a commonly used method to evaluate cuff pressure of the endotracheal tube after intubation. This study determined whether the size of the tracheal tube influenced the accuracy of pilot balloon palpation. A prospective observational analysis of 208 patients intubated with an endotracheal tube of internal diameter (ID) 6.0 or 8.0 was conducted. An anesthesiologist judged the cuff pressure by manual pilot balloon palpation, and then measured the cuff pressure with a pressure gauge. Cuff pressure exceeding 20–30 cmH
2
O was defined as false recognition. The intracuff pressure was significantly higher in ID 6.0 tube than in the ID 8.0 tube (41.9 ± 18.8 cmH
2
O vs. 30.3 ± 11.9 cmH
2
O,
p |
---|---|
ISSN: | 2045-2322 2045-2322 |
DOI: | 10.1038/s41598-023-32704-3 |