EXPERIMENTAL STUDY OF STABILITY AT CEMENT AND CEMENTLESS IMPLANTATION OF TRANSPEDICULAR SCREWS INTO VERTEBRAE WITH LOW MINERAL DENSITY OF BONE
The stability of cemented and noncemented transpedicular screw implantations in vertebra with low mineral bone density was studied in experiment. We analyzed the destabilization of screw under distraction load, attached along longitudal axis of the screw on universal servo-hydraulic test machine «Wa...
Gespeichert in:
Veröffentlicht in: | Travmatologii͡a︡ i ortopedii͡a︡ Rossii 2016-11, Vol.16 (2), p.97-101 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng ; rus |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The stability of cemented and noncemented transpedicular screw implantations in vertebra with low mineral bone density was studied in experiment. We analyzed the destabilization of screw under distraction load, attached along longitudal axis of the screw on universal servo-hydraulic test machine «Walter+bay ag» LFV-10-T50. The factors of bone-cement-metallic block rigidity of cemented screw was in 2,10-2,38 times more then in situation with traditional transpedicular screw. |
---|---|
ISSN: | 2311-2905 2542-0933 |
DOI: | 10.21823/2311-2905-2010-0-2-97-101 |