Testing and modeling HpGe detector with anti-Compton shield
Simulating with GEANT4 of anti-Compton shield containing eight BGO scintillators that works on anti-coincidence with HpGe detector was presented. The coefficient value of Compton tail suppression on 60Co spectrum was taken by simulating and from experiment. Comparing results of experimental data wit...
Gespeichert in:
Veröffentlicht in: | Eurasian journal of physics and functional materials 2021-11, Vol.4 (3), p.213-225 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Simulating with GEANT4 of anti-Compton shield containing eight BGO scintillators that works on anti-coincidence with HpGe detector was presented. The coefficient value of Compton tail suppression on 60Co spectrum was taken by simulating and from experiment. Comparing results of experimental data with simulating were presented. The influence of anti-coincidence inclusion on the detector efficiency was determined with simulation. |
---|---|
ISSN: | 2522-9869 2616-8537 |
DOI: | 10.29317/ejpfm.2020040303 |