Thresholds for oximetry alarms and target range in the NICU: an observational assessment based on likely oxygen tension and maturity
Continuous monitoring of SpO in the neonatal ICU is the standard of care. Changes in SpO exposure have been shown to markedly impact outcome, but limiting extreme episodes is an arduous task. Much more complicated than setting alarm policy, it is fraught with balancing alarm fatigue and compliance....
Gespeichert in:
Veröffentlicht in: | BMC pediatrics 2020-06, Vol.20 (1), p.317-8, Article 317 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Continuous monitoring of SpO
in the neonatal ICU is the standard of care. Changes in SpO
exposure have been shown to markedly impact outcome, but limiting extreme episodes is an arduous task. Much more complicated than setting alarm policy, it is fraught with balancing alarm fatigue and compliance. Information on optimum strategies is limited.
This is a retrospective observational study intended to describe the relative chance of normoxemia, and risks of hypoxemia and hyperoxemia at relevant SpO
levels in the neonatal ICU. The data, paired SpO
-PaO
and post-menstrual age, are from a single tertiary care unit. They reflect all infants receiving supplemental oxygen and mechanical ventilation during a 3-year period. The primary measures were the chance of normoxemia (PaO
50-80 mmHg), risks of severe hypoxemia (PaO
≤ 40 mmHg), and of severe hyperoxemia (PaO
≥ 100 mmHg) at relevant SpO
levels.
Neonates were categorized by postmenstrual age: 36 (n = 1031) weeks. From these infants, 26,162 SpO
-PaO
pairs were evaluated. The post-menstrual weeks (median and IQR) of the three groups were: 26 (24-28) n = 2603; 34 (33-35) n = 2501; and 38 (37-39) n = 21,058. The chance of normoxemia (65, 95%-CI 64-67%) was similar across the SpO
range of 88-95%, and independent of PMA. The increasing risk of severe hypoxemia became marked at a SpO
of 85% (25, 95%-CI 21-29%), and was independent of PMA. The risk of severe hyperoxemia was dependent on PMA. For infants 36 weeks at 96% SpO
(20, 95%-CI 17-22%).
The risk of hyperoxemia and hypoxemia increases exponentially as SpO
moves towards extremes. Postmenstrual age influences the threshold at which the risk of hyperoxemia became pronounced, but not the thresholds of hypoxemia or normoxemia. The thresholds at which a marked change in the risk of hyperoxemia and hypoxemia occur can be used to guide the setting of alarm thresholds. Optimal management of neonatal oxygen saturation must take into account concerns of alarm fatigue, staffing levels, and FiO
titration practices. |
---|---|
ISSN: | 1471-2431 1471-2431 |
DOI: | 10.1186/s12887-020-02225-3 |