Thresholds for oximetry alarms and target range in the NICU: an observational assessment based on likely oxygen tension and maturity

Continuous monitoring of SpO in the neonatal ICU is the standard of care. Changes in SpO exposure have been shown to markedly impact outcome, but limiting extreme episodes is an arduous task. Much more complicated than setting alarm policy, it is fraught with balancing alarm fatigue and compliance....

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:BMC pediatrics 2020-06, Vol.20 (1), p.317-8, Article 317
Hauptverfasser: Bachman, Thomas E, Iyer, Narayan P, Newth, Christopher J L, Ross, Patrick A, Khemani, Robinder G
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Continuous monitoring of SpO in the neonatal ICU is the standard of care. Changes in SpO exposure have been shown to markedly impact outcome, but limiting extreme episodes is an arduous task. Much more complicated than setting alarm policy, it is fraught with balancing alarm fatigue and compliance. Information on optimum strategies is limited. This is a retrospective observational study intended to describe the relative chance of normoxemia, and risks of hypoxemia and hyperoxemia at relevant SpO levels in the neonatal ICU. The data, paired SpO -PaO and post-menstrual age, are from a single tertiary care unit. They reflect all infants receiving supplemental oxygen and mechanical ventilation during a 3-year period. The primary measures were the chance of normoxemia (PaO 50-80 mmHg), risks of severe hypoxemia (PaO  ≤ 40 mmHg), and of severe hyperoxemia (PaO  ≥ 100 mmHg) at relevant SpO levels. Neonates were categorized by postmenstrual age:  36 (n = 1031) weeks. From these infants, 26,162 SpO -PaO pairs were evaluated. The post-menstrual weeks (median and IQR) of the three groups were: 26 (24-28) n = 2603; 34 (33-35) n = 2501; and 38 (37-39) n = 21,058. The chance of normoxemia (65, 95%-CI 64-67%) was similar across the SpO range of 88-95%, and independent of PMA. The increasing risk of severe hypoxemia became marked at a SpO of 85% (25, 95%-CI 21-29%), and was independent of PMA. The risk of severe hyperoxemia was dependent on PMA. For infants  36 weeks at 96% SpO (20, 95%-CI 17-22%). The risk of hyperoxemia and hypoxemia increases exponentially as SpO moves towards extremes. Postmenstrual age influences the threshold at which the risk of hyperoxemia became pronounced, but not the thresholds of hypoxemia or normoxemia. The thresholds at which a marked change in the risk of hyperoxemia and hypoxemia occur can be used to guide the setting of alarm thresholds. Optimal management of neonatal oxygen saturation must take into account concerns of alarm fatigue, staffing levels, and FiO titration practices.
ISSN:1471-2431
1471-2431
DOI:10.1186/s12887-020-02225-3