Leaf open time sinogram (LOTS): a novel approach for patient specific quality assurance of total marrow irradiation

There is no ideal detector-phantom combination to perform patient specific quality assurance (PSQA) for Total Marrow (TMI) and Lymphoid (TMLI) Irradiation plan. In this study, 3D dose reconstruction using mega voltage computed tomography detectors measured Leaf Open Time Sinogram (LOTS) was investig...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Radiation oncology (London, England) England), 2020-10, Vol.15 (1), p.1-236, Article 236
Hauptverfasser: Thiyagarajan, Rajesh, Sharma, Dayananda Shamurailatpam, Kaushik, Suryakant, Sawant, Mayur, Ganapathy, K., Nambi Raj, N. Arunai, Chilukuri, Srinivas, Sundar, Sham C., Patro, Kartikeswar Ch, Manikandan, Arjunan, Noufal, M. P., Sivaraman, Rangasamy, Easow, Jose, Jalali, Rakesh
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:There is no ideal detector-phantom combination to perform patient specific quality assurance (PSQA) for Total Marrow (TMI) and Lymphoid (TMLI) Irradiation plan. In this study, 3D dose reconstruction using mega voltage computed tomography detectors measured Leaf Open Time Sinogram (LOTS) was investigated for PSQA of TMI/TMLI patients in helical tomotherapy. The feasibility of this method was first validated for ten non-TMI/TMLI patients, by comparing reconstructed dose with (a) ion-chamber (IC) and helical detector array (ArcCheck) measurement and (b) planned dose distribution using 3Dγ analysis for 3%@3mm and dose to 98% (D 98% ) and 2% (D 2% ) of PTVs. Same comparison was extended for ten treatment plans from five TMI/TMLI patients. In all non-TMI/TMLI patients, reconstructed absolute dose was within ± 1.80% of planned and IC measurement. The planned dose distribution agreed with reconstructed and ArcCheck measured dose with mean (SD) 3Dγ of 98.70% (1.57%) and 2Dγ of 99.48% (0.81%). The deviation in D 98% and D 2% were within 1.71% and 4.10% respectively. In all 25 measurement locations from TMI/TMLI patients, planned and IC measured absolute dose agreed within ± 1.20%. Although sectorial fluence verification using ArcCHECK measurement for PTVs chest from the five upper body TMI/TMLI plans showed mean ± SD 2Dγ of 97.82% ± 1.27%, the reconstruction method resulted poor mean (SD) 3Dγ of 92.00% (± 5.83%), 64.80% (± 28.28%), 69.20% (± 30.46%), 60.80% (± 19.37%) and 73.2% (± 20.36%) for PTVs brain, chest, torso, limb and upper body respectively. The corresponding deviation in median D 98% and D 2% of all PTVs were 
ISSN:1748-717X
1748-717X
DOI:10.1186/s13014-020-01669-2