Probable edge defect in Acrysof single-piece intraocular lens

The purpose of this article is to report an edge defect in single-piece hydrophobic acrylic intraocular lens (IOL) observed during a routine phacoemulsification procedure. The chip was successfully removed intraoperatively with a McPherson's forceps. However, six months postoperatively patient...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Indian journal of ophthalmology 2006-12, Vol.54 (4), p.277-278
1. Verfasser: Shah, Urmi
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The purpose of this article is to report an edge defect in single-piece hydrophobic acrylic intraocular lens (IOL) observed during a routine phacoemulsification procedure. The chip was successfully removed intraoperatively with a McPherson's forceps. However, six months postoperatively patient complained of a disturbing glare due to posterior capsular folds originating from the site of IOL defect. Patient was advised, however refused to undergo Nd:Yag laser capsulotomy to relieve the symptoms. This case highlights the necessity to check each foldable IOL under the microscope before loading it into the cartridge, for possible edge defects.