Unfortunately Fortunate
An 88-year-old female presented with inappropriate implantable cardioverter-defibrillator shocks. Interrogation revealed lead noise sensed as ventricular fibrillation, leading to 11 shocks in 1 day. A defibrillation at the period of maximal vulnerability caused true ventricular fibrillation/ventricu...
Gespeichert in:
Veröffentlicht in: | JACC. Case reports 2020-02, Vol.2 (2), p.216-222 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An 88-year-old female presented with inappropriate implantable cardioverter-defibrillator shocks. Interrogation revealed lead noise sensed as ventricular fibrillation, leading to 11 shocks in 1 day. A defibrillation at the period of maximal vulnerability caused true ventricular fibrillation/ventricular tachycardia and additional shocks, which failed to terminate device-initiated tachyarrhythmia. The patient recovered spontaneously and the device was reprogrammed. (Level of Difficulty: Intermediate.)
[Display omitted]
An 88-year-old female presented with inappropriate implantable cardioverter-defibrillator shocks. Interrogation revealed lead noise sensed as… |
---|---|