Unfortunately Fortunate

An 88-year-old female presented with inappropriate implantable cardioverter-defibrillator shocks. Interrogation revealed lead noise sensed as ventricular fibrillation, leading to 11 shocks in 1 day. A defibrillation at the period of maximal vulnerability caused true ventricular fibrillation/ventricu...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:JACC. Case reports 2020-02, Vol.2 (2), p.216-222
Hauptverfasser: Ryaboy, Ilya V., Peele, Mark E.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:An 88-year-old female presented with inappropriate implantable cardioverter-defibrillator shocks. Interrogation revealed lead noise sensed as ventricular fibrillation, leading to 11 shocks in 1 day. A defibrillation at the period of maximal vulnerability caused true ventricular fibrillation/ventricular tachycardia and additional shocks, which failed to terminate device-initiated tachyarrhythmia. The patient recovered spontaneously and the device was reprogrammed. (Level of Difficulty: Intermediate.) [Display omitted] An 88-year-old female presented with inappropriate implantable cardioverter-defibrillator shocks. Interrogation revealed lead noise sensed as…