Evaluation of endothelial/Descemet membrane complex of eye bank donor corneas using enhanced depth imaging optical coherence tomography
We present a novel method for screening eye bank donor corneas using high definition optical coherence tomography (HD-OCT). This technology allows for the quantification of endothelial/Descemet membrane (En/DM) complex thickness ex vivo. Prospective interventional study. Fifty-two corneal grafts fro...
Gespeichert in:
Veröffentlicht in: | Clinical ophthalmology (Auckland, N.Z.) N.Z.), 2019-05, Vol.13, p.789-794 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We present a novel method for screening eye bank donor corneas using high definition optical coherence tomography (HD-OCT). This technology allows for the quantification of endothelial/Descemet membrane (En/DM) complex thickness ex vivo.
Prospective interventional study.
Fifty-two corneal grafts from 27 donors were included in this study. Twenty additional control eyes and 11 eyes with Fuchs' endothelial corneal dystrophy were also evaluated for comparison.
A custom built, high speed HD-OCT device (Envisu R2210, Bioptigen, Buffalo Grove, IL, USA) was used to obtain images, and custom-made graph-based segmentation software was used to automatically deconstruct corneal images into micro-layers. HD-OCT imaging was used to scan through the sealed sterile case of donor corneas stored in McCarey-Kaufman medium to image their En/DM complex through the center of the cornea.
This technology allowed for quantification of En/DM complex thickness in all donor corneas through the sealed sterile container used to transport graft tissue. Mean En/DM complex thickness of donor corneas was 17±4 μm. The difference between donor cornea En/DM thickness and that of control subjects (16±2 μm) was not statistically significant (
=0.3), suggesting that the transport container and media do not affect measurements. There was a significant difference between En/DM thickness of Fuchs' endothelial corneal dystrophy eyes (25±5 μm) and both donor corneas ( |
---|---|
ISSN: | 1177-5467 1177-5483 1177-5483 |
DOI: | 10.2147/OPTH.S185455 |