Multicentre experience with Optimus balloon-expandable cobalt-chromium stents in congenital heart disease interventions
To evaluate bare-metal Optimus and polytetrafluoroethylene (PTFE)-covered Optimus-CVS balloon-expandable, cobalt-chromium, hybrid cell-designed stents in congenital heart disease (CHD) interventions. Retrospective multicentre review of patients with CHD receiving Optimus stents. Stent mechanical beh...
Gespeichert in:
Veröffentlicht in: | Open heart 2023-01, Vol.10 (1), p.e002157 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To evaluate bare-metal Optimus and polytetrafluoroethylene (PTFE)-covered Optimus-CVS balloon-expandable, cobalt-chromium, hybrid cell-designed stents in congenital heart disease (CHD) interventions.
Retrospective multicentre review of patients with CHD receiving Optimus stents. Stent mechanical behaviour, clinical indications and outcomes were assessed.
183 stents (49.2% XXL/15-ZIG, 33.3% XL/12-ZIG, 17.5% L/9-ZIG) were implanted (98.9% success rate, 2.3% serious procedural complication rate) in 170 patients (57.6% men, 64.1% adults), median age 23.6 years (IQR, 15.2-39.2) and weight 63.5 kg (IQR, 47-75.7). Indications were right ventricular outflow tract stand-alone stenting or before revalvulation (62.4%), aortic coarctation treatment (15.3%), Fontan-circuit fenestration closure (12.4%) and miscellaneous (10%). 86/170 (50.6%) patients had PTFE-covered stenting (50% prophylactic). In 86/170 (50.6%) patients with stenotic lesions, median percentage of achieved stent expansion was 93.4% (IQR, 85.5%-97.7%), median gradient decreased from 28 mm Hg (IQR, 19-41) to 5 mm Hg (IQR, 1-9) (p |
---|---|
ISSN: | 2053-3624 2398-595X 2053-3624 |
DOI: | 10.1136/openhrt-2022-002157 |