Abstract We report a case of bead shedding from a cylindrical extensively porous-coated cementless femoral component with concomitant taper corrosion at the modular head-neck junction of a metal-on-polyethylene total hip prosthesis. The patient presented with chronic thigh pain 4 years after primary...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Arthroplasty today 2017-12, Vol.3 (4), p.215-219
Hauptverfasser: Edelstein, Adam I., MD, Phan, Duy L., MD, Sporer, Scott M., MD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract We report a case of bead shedding from a cylindrical extensively porous-coated cementless femoral component with concomitant taper corrosion at the modular head-neck junction of a metal-on-polyethylene total hip prosthesis. The patient presented with chronic thigh pain 4 years after primary total hip arthroplasty, and radiographs revealed significant osteolysis and metallic debris around the femoral stem. Intraoperatively, the patient had a grossly loose femoral component with debonding of sintered beads from the femoral stem, as well as evidence of taper corrosion. We identify a failure of a modern beaded femoral component in conjunction with taper corrosion.