Safe right bundle branch block pattern during permanent right ventricular pacing: a case report
We present the case of a 72-year-old male who underwent a permanent transvenous DDD pacemaker implantation for symptomatic complete heart block. The postoperative surface electrocardiogram (ECG) revealed a right bundle branch block (RBBB) pacing pattern with transition in lead V3 and left bundle bra...
Gespeichert in:
Veröffentlicht in: | Türk Kardiyoloji Derneği arşivi 2013-09, Vol.41 (6), p.522-525 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We present the case of a 72-year-old male who underwent a permanent transvenous DDD pacemaker implantation for symptomatic complete heart block. The postoperative surface electrocardiogram (ECG) revealed a right bundle branch block (RBBB) pacing pattern with transition in lead V3 and left bundle branch block (LBBB) in D1, suggesting a malpositioned ventricular lead in the left heart. Malposition of a lead was ruled out by chest X-ray and echocardiogram confirming apical right ventricular pacing. After a literature review, a surface ECG maneuver, by placing leads V1 and V2 one intercostal space lower, restored the usual patterns of paced LBBB pattern. We discuss the literature concerning safe-paced RBBB. |
---|---|