Reliability, repeatability, and accordance between three different corneal diagnostic imaging devices for evaluating the ocular surface

PurposeTo evaluate repeatability, reproducibility, and accordance between ocular surface measurements within three different imaging devices.MethodsWe performed an observational study on 66 healthy eyes. Tear meniscus height, non-invasive tear break-up time (NITBUT) and meibography were measured usi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Frontiers in medicine 2022-07, Vol.9, p.893688-893688
Hauptverfasser: Garcia-Terraza, Abril L., Jimenez-Collado, David, Sanchez-Sanoja, Francisco, Arteaga-Rivera, José Y., Morales Flores, Norma, Pérez-Solórzano, Sofía, Garfias, Yonathan, Graue-Hernández, Enrique O., Navas, Alejandro
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:PurposeTo evaluate repeatability, reproducibility, and accordance between ocular surface measurements within three different imaging devices.MethodsWe performed an observational study on 66 healthy eyes. Tear meniscus height, non-invasive tear break-up time (NITBUT) and meibography were measured using three corneal imaging devices: Keratograph 5M (Oculus, Wetzlar, Germany), Antares (Lumenis, Sidney, Australia), and LacryDiag (Quantel Medical, Cournon d'Auvergne, France). One-way ANOVAs with post hoc analyses were used to calculate accordance between the tear meniscus and NITBUT. Reproducibility was assessed through coefficients of variation and repeatability with intraclass correlation coefficients (ICC). Reliability of meibography classification was analyzed by calculating Fleiss' Kappa Index and presented in Venn diagrams.ResultsCoefficients of variation were high and differed greatly depending on the device and measurement. ICCs showed moderate reliability of NITBUT and tear meniscus height measurements. We observed discordance between measurements of tear meniscus height between the three devices, F2, 195 = 15.24, p < 0.01. Measurements performed with Antares were higher; 0.365 ± 0.0851, than those with Keratograph 5M and LacryDiag; 0.293 ± 0.0790 and 0.306 ± 0.0731. NITBUT also showed discordance between devices, F2, 111 = 13.152, p < 0.01. Measurements performed with LacryDiag were lower (10.4 ± 1.82) compared to those of Keratograph 5M (12.6 ± 4.01) and Antares (12.6 ± 4.21). Fleiss' Kappa showed a value of -0.00487 for upper lid and 0.128 for inferior lid Meibography classification, suggesting discrete to poor agreement between measurements.ConclusionDepending on the device used and parameter analyzed, measurements varied between each other, showing a difference in image processing.
ISSN:2296-858X
2296-858X
DOI:10.3389/fmed.2022.893688