Analytical evaluation of a BNP assay on the new point-of-care platform respons®IQ
a) Objectives: respons®IQ is a new point-of-care (POC) immunoassay platform utilizing evanescent field total internal reflection fluorescence (TIRF) detection and active microfluidics controlled by optical sensors. A B-type natriuretic peptide (BNP) assay was developed on this system. The objective...
Gespeichert in:
Veröffentlicht in: | Practical laboratory medicine 2015-08, Vol.2, p.15-21 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | a) Objectives: respons®IQ is a new point-of-care (POC) immunoassay platform utilizing evanescent field total internal reflection fluorescence (TIRF) detection and active microfluidics controlled by optical sensors. A B-type natriuretic peptide (BNP) assay was developed on this system. The objective was to show that the BNP test fulfils the basic requirements regarding analytical performance, storage stability of cartridges and correlation to reference systems to be used as a POC test.
b) Design and methods: Analytical sensitivity and imprecision were determined in 10 separate experiments over a period of one year. Cartridge storage stability at 4–7°C and 37°C was tested. The correlation of responsIQ whole blood measurements to a POC reference device and a laboratory analyzer was determined using 100 patient samples.
c) Results: Limit of detection (LOD) was 2.3±1.0pg/ml BNP and within-run coefficient of variation (within-run CV) was 4.8±1.4% down to a concentration of |
---|---|
ISSN: | 2352-5517 2352-5517 |
DOI: | 10.1016/j.plabm.2015.04.002 |