False aminoglycoside resistance in Enterobacterales and non-fermenters by an automated testing system: a descriptive study

Antimicrobial sensitivity data are important to guide antimicrobial therapy. In microbiological laboratories, routine sensitivity measurements are typically performed with automated testing systems such as VITEK2 and Phoenix. Using data from the Dutch national surveillance system for antimicrobial r...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Microbiology spectrum 2023-12, Vol.11 (6), p.e0309323-e0309323
Hauptverfasser: Klein Klouwenberg, P M C, Altorf-van der Kuil, W, van Griethuysen, A J, Hendriks, M, Kuijper, E J, Notermans, D W, Schoffelen, A F
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Antimicrobial sensitivity data are important to guide antimicrobial therapy. In microbiological laboratories, routine sensitivity measurements are typically performed with automated testing systems such as VITEK2 and Phoenix. Using data from the Dutch national surveillance system for antimicrobial resistance over a 6-year period, we found that the measured minimum inhibitory concentrations for aminoglycosides in Enterobacterales and non-fermenters were too high for the Phoenix system. In addition, we observed a yearly increase in resistance for several species measured by Phoenix. These findings might have consequences for clinical treatment of patients with sepsis.
ISSN:2165-0497
2165-0497
DOI:10.1128/spectrum.03093-23