Assessment of the reliability of a non-invasive elbow valgus laxity measurement device
Purpose The purpose of this study was to assess the reliability of a new objective measurement tool to measure the valgus stress laxity of the ulnar collateral ligament (UCL) of the elbow, the “Elbow Tester”. The anterior oblique portion of the ulnar collateral ligament (UCL) of the elbow is the pri...
Gespeichert in:
Veröffentlicht in: | Journal of Experimental Orthopaedics 2020-09, Vol.7 (1), p.74-74, Article 74 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
The purpose of this study was to assess the reliability of a new objective measurement tool to measure the valgus stress laxity of the ulnar collateral ligament (UCL) of the elbow, the “Elbow Tester”. The anterior oblique portion of the ulnar collateral ligament (UCL) of the elbow is the primary static restraint to valgus stress during the overhead throwing motion. The main research question was if the “Elbow Tester” that we have developed was reliable and reproducible for further use in research and daily practice.
Methods
Three different examiners tested both elbows of 11 volunteers for UCL laxity. Each elbow was tested 5 times using a standard 2 Nm valgus load, and 3 times using a manual maximum valgus load. One examiner tested the volunteers again 1 week later. The outcomes of elbow valgus laxity were compared between examiners. The intraobserver reliability was assessed using an intraclass correlation coefficient (ICC) and interobserver reliability was also assessed with a mixed model repeated ANOVA test.
Results
The device demonstrated a high level of intraobserver reliability with both the 2 Nm valgus force and manual maximum valgus force, using a minimum of three trials as determined by an ICC > 0.9 for all examiners. The interobserver reliability was moderate using the 2 Nm valgus load with an ICC value of 0.72 and significant different outcomes of elbow valgus laxity amongst examiners (
p
0.53).
Conclusion
The noninvasive valgus elbow tester demonstrates high interobserver and intraobserver reliability using manual maximum valgus force and can be used for further research and daily practice. |
---|---|