FT-IR Spectroscopy for the Detection of Diethylene Glycol (DEG) Contaminant in Glycerin-Based Pharmaceutical Products and Food Supplements

Identification and determination of diethylene glycol (DEG) in glycerin-based products was successfully achieved using FT-IR spectroscopy. Studied samples included 0.5% to 20% by mass DEGspiked into cough syrup, two paracetamol syrup formulations, and two food supplements. The characteristic DEGwave...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Acta chimica Slovenica 2020-06, Vol.67 (2), p.530-536
Hauptverfasser: Hammoudeh, Ayman Y, Obeidat, Safwan M, Abboushi, Eman Kh, Mahmoud, Amal M
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Identification and determination of diethylene glycol (DEG) in glycerin-based products was successfully achieved using FT-IR spectroscopy. Studied samples included 0.5% to 20% by mass DEGspiked into cough syrup, two paracetamol syrup formulations, and two food supplements. The characteristic DEGwavenumbers at 881 cm-1 and 1083 cm-1 were used for its quantitative determination in the studied samples. A very good accuracy in determining the DEG fraction was achieved with a mean error% of ±2.02% to ±7.69% upon using the corrected absorbance at 881 cm-1. The corrected absorbance at 1083 cm-1 band was used in the case of paracetamol formulations and resulted in a mean error% ranging from ±2.50% to ±10.28%. The values of limit of detection of the current method ranged from 0.051% to 0.068% DEG for all studied samples.
ISSN:1580-3155
1318-0207
1580-3155
DOI:10.17344/ACSI.2019.5553