PERSENTASE DOSIS KEDALAMAN (PERCENTAGE DEPTH DOSE) DENGAN PESAWAT TELETERAPI COBALT-60 FCC 8000 F DI RSUP SANGLAH DENPASAR
Successful implementation of treatment using Cobalt-60 8000 F FCC, strongly influenced the precision and accuracy of the implementation is largely determined by the data PDD. Based on the research the value of PDD at a depth of 0 cm with a broad range of the field is at 39.7% - 48.5%, while the maxi...
Gespeichert in:
Veröffentlicht in: | Buletin Fisika (Online) 2015-02, Vol.16 (1), p.1-6 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Successful implementation of treatment using Cobalt-60 8000 F FCC, strongly influenced the precision and accuracy of the implementation is largely determined by the data PDD. Based on the research the value of PDD at a depth of 0 cm with a broad range of the field is at 39.7% - 48.5%, while the maximum value reached 100% PDD found at a depth of 0.5 cm. The amount of deviation between the results of a study of the data and BJR Sanglah reached 5% at a depth of 8 cm by 6 cm wide field x 6 cm and 8 cm x 8 cm, and the smallest deviation value of 0.1% found in a wide field 6 cm x 6 cm with depth 7 cm. keywords: cobalt-60, the percentage depth dose, dispersion values |
---|---|