Aqueductal CSF stroke volume measurements may drive management of shunted idiopathic normal pressure hydrocephalus patients

CSF shunting with adjustable valve is the treatment of idiopathic normal pressure hydrocephalus. The opening pressure valve setting is left to the neurosurgeon’s experience. Aqueductal CSF stroke volume by phase-contrast magnetic resonance measures the CSF passing through the Sylvian aqueduct and it...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Scientific reports 2021-03, Vol.11 (1), p.7095-7095, Article 7095
Hauptverfasser: Scollato, Antonio, Caini, Saverio, Angelini, Lucia, Lastrucci, Giancarlo, Di Lorenzo, Nicola, Porfirio, Berardino, Gallina, Pasquale
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:CSF shunting with adjustable valve is the treatment of idiopathic normal pressure hydrocephalus. The opening pressure valve setting is left to the neurosurgeon’s experience. Aqueductal CSF stroke volume by phase-contrast magnetic resonance measures the CSF passing through the Sylvian aqueduct and it changes with intracranial hydrodynamics. We sought to identify a window of stroke volume differences associated with the best clinical outcome and lowest rate of complications. The records of 69 patients were reviewed. At every clinical check, stroke volume, opening pressure valve, clinical outcome, and CSF overdrainage were analyzed. The correlation between stroke volume differences and negative outcome was also analyzed. The median follow-up was 2.3 years (range 0.3–10.4 years). The odds of negative outcome between two consecutive checks significantly increased by 16% (95%CI 4–28%, p  = 0.006). Taking the lowest risk group as reference, the odds ratio of negative outcome was 1.16 (95%CI 0.51–2.63, p  = 0.726) for SV differences less than − 37.6 µL, while it was 1.96 (95%CI 0.97–3.98, p  = 0.062) for stroke volume changes above + 23.1 µL. Maintaining stroke volume values within a definite range might help maximize clinical benefit and avoid the risk of CSF overdrainage.
ISSN:2045-2322
2045-2322
DOI:10.1038/s41598-021-86350-8