Novel Tools for Single Comparative and Unified Evaluation of Qualitative and Quantitative Bioassays: SS/PV-ROC and SS-J/PV-PSI Index-ROC Curves with Integrated Concentration Distributions, and SS-J/PV-PSI Index Cut-Off Diagrams

This investigation is both a study of potential non-invasive diagnostic approaches for the bladder cancer biomarker UBC Rapid test and a study including novel comparative methods for bioassay evaluation and comparison that uses bladder cancer as a useful example. The objective of the paper is not to...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Diagnostics (Basel) 2024-05, Vol.14 (9), p.951
1. Verfasser: Oehr, Peter
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This investigation is both a study of potential non-invasive diagnostic approaches for the bladder cancer biomarker UBC Rapid test and a study including novel comparative methods for bioassay evaluation and comparison that uses bladder cancer as a useful example. The objective of the paper is not to investigate specific data. It is used only for demonstration, partially to compare ROC methodologies and also to show how both sensitivity/specificity and predictive values can be used in clinical diagnostics and decision making. This study includes ROC curves with integrated cut-off distribution curves for a comparison of sensitivity/specificity (SS) and positive/negative predictive values (PPV/NPV or PV), as well as SS-J index/PV-PSI index-ROC curves and SS-J/PV-PSI index cut-off diagrams (J = Youden, PSI = Predictive Summary Index) for the unified direct comparison of SS-J/PV results achieved via quantitative and/or qualitative bioassays and an identification of optimal separate or unified index cut-off points. According to the routine diagnostics, there were 91 patients with confirmed bladder cancer and 1152 patients with no evidence of bladder cancer, leading to a prevalence value of 0.073. This study performed a quantitative investigation of used-up test cassettes from the visual UBC Rapid qualitative point-of-care assay, which had already been applied in routine diagnostics. Using a photometric reader, quantitative data could also be obtained from the test line of the used cassettes. Interrelations between SS and PV values were evaluated using cumulative distribution analysis (CAD), SS/PV-ROC curves, SS-J/PV-PSI index-ROC curves, and the SS-J/PV-PSI index cut-off diagram. The maximum unified SS-J/PV-PSI index value and its corresponding cut-off value were determined and calculated with the SS-J/PV-PSI index cut-off diagram. The use of SS/PV-ROC curves with integrated cut-off concentration distribution curves provides improved diagnostic information compared to "traditional" ROC curves. The threshold distributions integrated as curves into SS/PV-ROC curves and SS-J/PV-PSI index-ROC curves run in opposite directions. In contrast to the SS-ROC curves, the PV-ROC and the novel PV-PSI index-ROC curves had neither an area under the curve (AUC) nor a range from 0% to 100%. The cut-off level of the qualitative assay was 7.5 µg/L, with a sensitivity of 65.9% and a specificity of 63.3%, and the PPV was 12.4% and the NPV was 95.9%, at a threshold value of 12.5 µg/L
ISSN:2075-4418
2075-4418
DOI:10.3390/diagnostics14090951