Validation and reliability of a prototype orthodontic bracket debonding device equipped with force-sensitive resistor (FSR): a novel method of measuring orthodontic bracket debonding force in vivo

Background To introduce an orthodontic bracket debonding device capable of measuring debonding force clinically by a novel sensor mechanism Materials and method A prototype orthodontic debonding device was constructed utilizing a lift-off debonding instrument (LODI) and force-sensitive resistor (FSR...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Progress in Orthodontics 2019-07, Vol.20 (1), p.26-8, Article 26
Hauptverfasser: Ahmed, Tamzid, Rahman, Norma Ab, Alam, Mohammad Khursheed
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Background To introduce an orthodontic bracket debonding device capable of measuring debonding force clinically by a novel sensor mechanism Materials and method A prototype orthodontic debonding device was constructed utilizing a lift-off debonding instrument (LODI) and force-sensitive resistor (FSR). For data interpretation, the force sensor was equipped with a microcontroller and C++ programming software running on a computer. Ninety-nine (99) 0.022-in. conventional metallic brackets were bonded to premolar teeth in vitro by a single clinician applying the same adhesive and bonding technique. For validation, the mean debonding force measured by the prototype debonding device ( n  = 30) and the universal testing machine ( n  = 30) was compared. Both intra- and inter-examiner reliability tests were done by holding and operating the device in a standardized manner. Following debonding by the prototype device, the bracket failure pattern was evaluated ( n  = 30) by adhesive remnant index (ARI) under the stereomicroscope at × 30 magnification. Statistical analysis included independent samples t test for validation and intraclass correlation coefficient (ICC) with a 95% confidence interval for both intra- and inter-examiner reliability. Results Mean orthodontic bracket debonding force measured by the prototype device (9.36 ± 1.65 N) and the universal testing machine (10.43 ± 2.71 N) was not significantly different ( p  
ISSN:2196-1042
1723-7785
2196-1042
DOI:10.1186/s40510-019-0277-x