Evaluation of an articulated passive ankle-foot prosthesis
Current ankle prostheses for people with unilateral transtibial amputation (TTA) or transfemoral amputation (TFA) are unable to mimic able-bodied performance during daily activities. A new mechanical ankle-foot prosthesis was developed to further optimise the gait of people with a lower-limb amputat...
Gespeichert in:
Veröffentlicht in: | Biomedical engineering online 2022-04, Vol.21 (1), p.28-28, Article 28 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Current ankle prostheses for people with unilateral transtibial amputation (TTA) or transfemoral amputation (TFA) are unable to mimic able-bodied performance during daily activities. A new mechanical ankle-foot prosthesis was developed to further optimise the gait of people with a lower-limb amputation. This study aimed to evaluate the Talaris Demonstrator (TD) during daily activities by means of performance-related, physiological and subjective outcome measures.
Forty-two participants completed a protocol assessing performance and functional mobility with their current prosthesis and the TD. The protocol comprised the L-test, 2 min of stair climbing, 2 min of inclined treadmill walking, 6 min of treadmill walking at 3 different speeds in consecutive blocks of 2 min, and a 3-m Backward Walk test (3mBWT). Heart rate was measured during each task, and oxygen uptake was collected during all tasks except for the L-test and 3mBWT. Time of execution was recorded on the L-test and 3mBWT, and the rate of perceived exertion (score = 6-20), fatigue and comfort (score = 0-100) were assessed after each task. Paired sample t-tests and Wilcoxon Signed-rank tests were performed to compare outcomes between prosthetic devices. Benjamini-Hochberg corrections were applied to control for multiple comparisons with a level of significance set at α = 0.05.
Subjects with a TTA (N = 28) were faster with their current prosthesis compared to the TD on the L-test and 3mBWT (p = 0.005). In participants with a TFA (N = 14), we observed a tendency towards a higher heart rate during the L-test and towards increased comfort during inclined walking, with the TD compared to the participants' current prosthetic device (0.05 |
---|---|