Development and Validation of Novel HPLC Methods for Quantitative Determination of Vitamin D3 in Tablet Dosage Form

In the present work, an efficient isocratic HPLC method was developed for the precise and accurate estimation of vitamin D in tablet form. The chromatographic conditions comprised an L3 silica column (5 µm in particle size, 4.6 mm × 250 mm) with a mobile phase n-hexane/ethyl acetate (85:15 / ) with...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pharmaceuticals (Basel, Switzerland) Switzerland), 2024-04, Vol.17 (4), p.505
Hauptverfasser: Gohar, Muhammad Saqib, Rahman, Taj Ur, Bahadur, Ali, Ali, Ashraf, Alharthi, Sarah, Al-Shaalan, Nora Hamad
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In the present work, an efficient isocratic HPLC method was developed for the precise and accurate estimation of vitamin D in tablet form. The chromatographic conditions comprised an L3 silica column (5 µm in particle size, 4.6 mm × 250 mm) with a mobile phase n-hexane/ethyl acetate (85:15 / ) with a flow rate of 2.0 mL/min and a detection wavelength of 292 nm. The new methodology was validated for accuracy, precision, specificity, robustness, and quantification limits according to an official monograph of USP/BP and ICH guidelines. The peak areas of the six replicates of the homogeneous sample were recorded. The mean value obtained was 67,301, and the relative standard deviation (RSD) was 0.1741. The linearity and range were in the acceptable bounds, i.e., 0.999, which was calculated using regression line analysis. The results show that the method is truly acceptable as the RSD, as the flow rate was 0.81%, while for the mobile phase composition, it was 0.72%, which lies in the acceptable range. The limit of detection (LOD) and the limit of quantification (LOQ) values were 0.0539 µg/mL and 0.1633 µg/mL, respectively. The % RSD of the intra and inter-day precision of the method was deemed acceptable according to the international commission for harmonization guidelines. The developed method has potential to be used for the detection and quantification of vitamin D during routine analysis for tablets in dosage form.
ISSN:1424-8247
1424-8247
DOI:10.3390/ph17040505