Wearable in-ear pulse oximetry validly measures oxygen saturation between 70% and 100%: A prospective agreement study

Objectives Postoperative monitoring outside intensive and post-anaesthesia care units is seldom, partly due to lack of suitable and approved systems. We therefore aim to validate the oxygen saturation (SpO2) and pulse rate measurement of the in-ear sensor c-med° alpha with a reference pulse oximeter...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Digital health 2023-01, Vol.9, p.20552076231211169-20552076231211169
Hauptverfasser: Bubb, Catherina AB, Weber, Michael, Kretsch, Nadine, Heim, Ralph, Zellhuber, Incinur, Schmid, Sebastian, Kagerbauer, Simone M, Kreuzer, Johannes, Schaller, Stefan J, Blobner, Manfred, Jungwirth, Bettina
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Objectives Postoperative monitoring outside intensive and post-anaesthesia care units is seldom, partly due to lack of suitable and approved systems. We therefore aim to validate the oxygen saturation (SpO2) and pulse rate measurement of the in-ear sensor c-med° alpha with a reference pulse oximeter. Methods This prospective agreement study was conducted in 12 healthy (ASA 1) adult (18–50 years) volunteers according to the EN ISO 80601-2-61. The sitting volunteers were equipped with the finger pulse oximeter Rad-5 and two c-med° alpha sensors in each ear. The inspiratory oxygen content was reduced via a tight-fitting breathing mask to achieve five defined plateaus with stable SpO2 between 99% and 70%. The deviation of the SpO2 and pulse rate measurements of the c-med° alpha from those of the Rad-5 was calculated using the mean square error (Arms). Bias and limits of agreement between both devices were calculated using the Bland-Altman technique. The precision was compared based on the repeatability coefficients. Results The c-med° alpha measured SpO2 had an Arms = 1.9% relative to the Rad-5, a non-significant bias (−0.1% (-0.2% to 0.0%)), levels of agreement from −4.0% to 3.8%, and the same repeatability coefficient (0.8% vs. 0.8%). The c-med° alpha measured pulse rate did not deviate from the one measured with the certified finger pulse oximeter (bias: 0.1 min−1 (0 to 0.1 min−1), level of agreement: −3.6 to 3.7 min−1, Arms: 1.8 min−1). Conclusions The c-med° alpha fulfils the EN ISO 80601-2-61 standard and is sufficiently accurate for measuring SpO2 and pulse rate in healthy adults at rest. Trial registration EUDAMED No. CIV-21-03-036033