Development of a Sequential Injection Analysis System for the Determination of Saccharin

Saccharin is a powerfully sweet nonnutritive sweetener that has been approved for food-processing applications within the range of 100-1200 mg/kg. A simple, rapid, and cost-effective sequential injection analysis (SIA) technique was developed to determine the saccharin level. This method is based on...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Sensors (Basel, Switzerland) Switzerland), 2017-12, Vol.17 (12), p.2891
Hauptverfasser: Wibowotomo, Budi, Eun, Jong-Bang, Rhee, Jong Il
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Saccharin is a powerfully sweet nonnutritive sweetener that has been approved for food-processing applications within the range of 100-1200 mg/kg. A simple, rapid, and cost-effective sequential injection analysis (SIA) technique was developed to determine the saccharin level. This method is based on the reaction of saccharin with p-chloranil in an ethanol medium with a hydrogen peroxide (H₂O₂) acceleration, and the resultant violet-red compound was detected using a UV-Vis spectrophotometer at λ = 420 nm. To ascertain the optimal conditions for the SIA system, several parameters were investigated, including buffer flow rate and volume, p-chloranil concentration, and reactant volumes (saccharin, p-chloranil, and H₂O₂). The optimum setup of the SIA system was achieved with a buffer flow rate, buffer volume, and draw-up time of 1.2 mL/min, 2900 µL, and ~145 s, respectively. The optimal p-chloranil concentration is 30 mM, and the best reactant volumes, presented in an ordered sequence, are as follows: 30 µL of H₂O₂, 450 µL of saccharin, and 150 µL of p-chloranil. The optimized SIA configuration produced a good linear calibration curve with a correlation coefficient (R² = 0.9812) in the concentration range of 20-140 mg/L and with a detection limit of 19.69 mg/L. Analytical applications in different food categories also showed acceptable recovery values in the range of 93.1-111.5%. This simple and rapid SIA system offers great feasibility for the saccharin quality control in food-product processing.
ISSN:1424-8220
1424-8220
DOI:10.3390/s17122891