Clinical evaluation of [123I]FP-CIT SPECT scans on the novel brain-dedicated InSPira HD SPECT system: a head-to-head comparison

The InSPira HD system, a novel brain-dedicated SPECT scanner, allows for imaging with a high spatial resolution. Here, we tested whether this scanner can be used to image the dopamine transporter adequately. Therefore, striatal phantom and patient data acquired on the InSPira were compared head-to-h...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:EJNMMI Research 2018-08, Vol.8 (1), p.85-6, Article 85
Hauptverfasser: Adriaanse, Sofie M., de Wit, Tim C., Stam, Mette, Verwer, Eline, de Bruin, Kora M., Booij, Jan
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The InSPira HD system, a novel brain-dedicated SPECT scanner, allows for imaging with a high spatial resolution. Here, we tested whether this scanner can be used to image the dopamine transporter adequately. Therefore, striatal phantom and patient data acquired on the InSPira were compared head-to-head with the well-validated brain-dedicated NeuroFocus system. A striatal phantom filled with [123I] and 14 subjects (after [123I]FP-CIT injection) were scanned on both systems. [123I]FP-CIT SPECT scans were visually assessed. Striatal binding ratios were calculated automatically using the software package BRASS. Striatal phantom and patient data showed strong correlations with respect to striatal ratios ( R  = 0.99 and R  = 0.92; p  
ISSN:2191-219X
2191-219X
DOI:10.1186/s13550-018-0436-y