Comparison of intraocular pressure as measured by three different non-contact tonometers and goldmann applanation tonometer for non-glaucomatous subjects

To compare the measurement of intraocular pressure (IOP) among the three different non-contact tonometers (NCT) and the Goldmann applanation tonometer (GAT) for non-glaucomatous subjects. In 52 eyes of 52 non-glaucomatous subjects, IOP was measured sequentially with the Canon TX-20P, the Nidek NT-53...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:BMC ophthalmology 2017-11, Vol.17 (1), p.199-199, Article 199
Hauptverfasser: Bang, Seung Pil, Lee, Chong Eun, Kim, Yu Cheol
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:To compare the measurement of intraocular pressure (IOP) among the three different non-contact tonometers (NCT) and the Goldmann applanation tonometer (GAT) for non-glaucomatous subjects. In 52 eyes of 52 non-glaucomatous subjects, IOP was measured sequentially with the Canon TX-20P, the Nidek NT-530P, the Topcon CT-1P, and the GAT at the same time. We evaluated the IOP-measurement agreement among the tonometers as well as the factors affecting the measurements. A significant positive correlation was shown between the IOP values obtained with GAT and each NCT. The Canon TX-20P showed statistically the most significant agreement with the GAT (ICC 0.906, 95% CI 0.837-0.946). In an analysis of the Bland-Altman plots, the Canon TX-20P also showed the largest mean bias (1.38 mmHg) but the narrowest limits of agreement (LoA) (95% LoA; ± 3.43 mmHg). The Topcon CT-1P showed the smallest mean bias (0.48 mmHg) but the widest LoA (95% LoA; ± 4.16 mmHg). The Topcon CT-1P and Nidek NT-530P both showed a significantly positive correlation between increasing central corneal thickness (CCT) and increasing IOP. There was a statistically significant correlation between each of the three different NCT and the GAT measurements. IOP measured with the Canon TX-20P and Topcon CT-1P tended to be higher, and with the Nidek NT-530P lower, than with the GAT. Practitioners need to know the properties of their own NCTs and their respective measurement tendencies.
ISSN:1471-2415
1471-2415
DOI:10.1186/s12886-017-0593-1