When the lead extractor becomes the assistant for new lead implant

With the increasing number of cardiovascular implantable electronic device upgrade and vein obstruction caused by previous leads, it is important to have alternative techniques to upgrade the device with the maintenance of functioning leads. We report an 83-year old male with 13-year old one-lead du...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Indian pacing and electrophysiology journal 2021-01, Vol.21 (1), p.59-61
Hauptverfasser: Nunes-Ferreira, Afonso, Bernardes, Ana, Pinto, Fausto J., de Sousa, João, Marques, Pedro
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:With the increasing number of cardiovascular implantable electronic device upgrade and vein obstruction caused by previous leads, it is important to have alternative techniques to upgrade the device with the maintenance of functioning leads. We report an 83-year old male with 13-year old one-lead dual-chamber pacemaker, ischemic cardiac disease and pre-dialytic chronic kidney disease submitted to an upgrade to cardiac resynchronization therapy. A sub-occlusion in the transition of left brachiocephalic vein and the superior vena cava was documented. Re-permeabilization was only achieved with a TightRail™ rotating dilator sheath over a guidewire with successful left ventricle lead implant.
ISSN:0972-6292
0972-6292
DOI:10.1016/j.ipej.2020.09.006