Clinical effectiveness and cost-effectiveness of depth of anaesthesia monitoring (E-Entropy, Bispectral Index and Narcotrend): a systematic review and economic evaluation
It is important that the level of general anaesthesia (GA) is appropriate for the individual patient undergoing surgery. If anaesthesia is deeper than required to keep a patient unconscious, there might be increased risk of anaesthetic-related morbidity, such as postoperative nausea, vomiting and co...
Gespeichert in:
Veröffentlicht in: | Health technology assessment (Winchester, England) England), 2013-08, Vol.17 (34), p.1-264 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | It is important that the level of general anaesthesia (GA) is appropriate for the individual patient undergoing surgery. If anaesthesia is deeper than required to keep a patient unconscious, there might be increased risk of anaesthetic-related morbidity, such as postoperative nausea, vomiting and cognitive dysfunction. This may also prolong recovery times, potentially increasing health-care costs. If anaesthesia is too light, patients may not be fully unconscious and could be at risk of intraoperative awareness.
The objective of this report is to assess the clinical effectiveness and cost-effectiveness of Bispectral Index (BIS), E-Entropy and Narcotrend technologies, each compared with standard clinical monitoring, to monitor the depth of anaesthesia in surgical patients undergoing GA.
A search strategy was developed and run on a number of bibliographic electronic databases including MEDLINE, EMBASE, The Cochrane Library and the Health Technology Assessment (HTA) database. For the systematic review of patient outcomes, databases were searched from the beginning of 2009 to November 2011 for studies of BIS (and then updated in February 2012), and from 1995 to November 2011 (and then updated in February 2012) for studies of E-Entropy and Narcotrend. For the systematic review of cost-effectiveness, searches were from database inception to November 2011 (an update search was performed in February 2012).
The systematic review of patient outcomes followed standard methodology for evidence synthesis. A decision-analytic model was developed to assess the cost-effectiveness of depth of anaesthesia monitoring compared with standard clinical observation. A simple decision tree was developed, which accounted for patients' risk of experiencing short-term anaesthetic-related complications in addition to risk of experiencing intraoperative awareness.
Twenty-two randomised controlled trials comparing BIS, E-Entropy and Narcotrend with standard clinical monitoring were included in the systematic review of patient outcomes, alongside evidence from a recent Cochrane review. Six trials of patients classified with risk factors for intraoperative awareness were combined in a fixed-effect meta-analysis. The overall pooled Peto's odds ratio was 0.45 (95% confidence interval 0.25 to 0.81) in favour of BIS. However, there was statistically significant heterogeneity. The base-case cost per quality-adjusted life-year (QALY) for BIS compared with standard clinical monitoring ranged fro |
---|---|
ISSN: | 1366-5278 2046-4924 |
DOI: | 10.3310/hta17340 |